35-5 35-5T 35-5P 35-5TP 56-4 56-4P  

 

Electrical Specifications

 

Output ranges:

 

 

 

Range 1

 

0-35 V, 0-3 A

 

Range 2

 

0-15 V, 0-5 A

 

Range 3

 

0-35 V, 0-500.0 mA

 

Outputs

 

3

 

Output power

 

215 W

 

Interface (GPIB/RS-232/USB)

 

No

 

Voltage setting

 

By floating point numeric entry or rotary jog wheel; resolution 1 mV

 

Current setting

 

By floating point numeric entry or rotary jog wheel; resolution 1 mV or 0.1 mA depending on range

 

Setting accuracy

 

Voltage - 0.03% +/- 5 mV. Current 0.2% +/- 5 mA, 0.5 mA

 

Output mode

 

Operation in constant voltage or constant current modes with automatice cross-over and mode indication by LEDs

 

DC output switch

 

Sets output voltage and current levels to zero when Off

 

Output terminals

 

4mm terminals on 19mm (0.75") spacing

 

Load regulation

 

Voltage: <0.01% + 2mV               
Current: <0.01% + 250mA; <0.01% + 50mA on 500mA range (measured at output terminals using remote sense)

 

Line regulation

 

Voltage: <0.01% + 2 mV for 10% line change           
Current: <0.01% + 250mA; <0.01% + 50mA on 500mA range

 

Ripple and Noise

 

Typically <0.35mV rms 2m V p-p CV mode, and <0.2mA rms, <20mA rms (500 mA range) CI mode

 

Transient response

 

50ms to within 15 mV of set level for a change in load current from full load to half load or vice versa

 

Temperature coefficient

 

< +/- (50ppm + 0.5 mV)/oC (voltage)

 

Remote sense

 

Eliminates up to 0.5 V drop per lead. Remote sense operation selected from front panel and indicated by LED

 

Sense terminals

 

Recessed sprung sockets for direct insertion of wires. Duplicated on rear terminal block (P versions only)

 

General Specifications

 

Power requirements

 

115 V or 230 V +/- 10% (adjustable internally, option HV for factory set 230 VAC input), 50/60 Hz. Installation Category ll

 

Operating range

 

5oC to 40oC, 20% to 80% RH

 

Storage range

 

-40oC to 70oC

 

Dimensions (HxWxD)

 

6.3 x 11 x 11.4"/160 x 280 x 290 mm

 

Weight

 

23.1 lbs/10.5 kg

 

Benchtop Operation

 

Folding legs are incorporated that can be used to angle the front panel upwards when required

 

Rack mount operation

 

19 inch 4U mount for up to three single output units or one triple plus one single; Blanking plates available for un-used sections

 

Warranty

 

3 years

 

Output Protection and Metering

 

Output protection

 

Output will withstand forward voltages of up to 20 V above rated output voltage. Reverse protection by diode clamp for current up to 3 A

 

Fault condition trip

 

The output will be shut down if any of the four trip conditions listed below occur. In addition to the output eing set Off, an isolated rear panel signal is also activated.

 

Over voltage (OVP)

 

Settable 1V to 40 V in 0.1 V steps

 

Over current (OCP)

 

Settable 0.1 A to 5.5 A in 0.01 A steps

 

Over temperature

 

Monitors internal temperature rise to protect against excess ambient temperature or blocked ventilation slots

 

Sense Error

 

Monitors the voltage between the remote sense terminals and output terminals to protect against mis-wiring

 

Trip output signal

 

Isolated open-collector output signal on rear panel

 

Meter resolution and accuracy:

 

Voltage (CI mode):

 

Resolution 10 mV     Accuracy +/- (0.1% of reading + 10 mV)

 

Current (CV mode):

 

Resolution 0.001 A; 0.1 A on 500 mA range   Accuracy +/- (0.2% + 0.005 A); +/- (0.2% + 0.5 mA) on 500 mA range

 

V x A

 

Resolution 0.01 W; 0.001 W on 500 mA range            Accuracy +/- (0.3% + 0.05 W); +/- (0.3% +0 0.005 W) on 500 mA range

 

Store/Recall Settings

 

Number of stores

 

30 plus power-down store

 

Memory type

 

Non-volatile using EEPROM

 

Parameters stored

 

Range, Set volts, Set current, OVP, OCP

 

Recall system

 

Settings are previewed onto the displays before being actioned

 

Bus Interfaces (P Suffix versions)

 

USB

 

Standard USB hardware connection  Supplied with device driver for Win 98 or above. Operates as a virtual COM port.

 

RS-232

 

Variable baud rate 19,200 max. Single instrument or Addressable RS-232

 

GPIB

 

Conforming with IEEE-488.1 and IEEE-488.2 (N.B. All three interfaces incorporate full control, readback and status reporting)

 

BUS type selection

 

From front panel (GPIB/RS-232/USB)

 

Address selection

 

From front panel (1 to 31)

 

Baud selection

 

RS-232 only. Fron front panel (600 to 19200 baud)

 

Setting resolution

 

Voltage - 1mV, Current 0.1 mA (0.1 mA on 500 mA range)

 

Accuracy

 

See specifications under Outputs and Metering

 

Remote control response time:

 

Interface

 

Typically < 80 ms

 

Output voltage

 

Response time varies with range and load conditions. Typical time to settle to within 1% of the total excursion on a 35 V / 3 A range with full load is <25 ms. With no load it is <7 ms for an upward charge and < 600 ms for downward.

 

Auxiliary Output-XDL 35-5T and  TP

 

Output voltage

 

Switchable 2.7 V, 3.3 V or 5.0 V. Accuracy better than +/- 5%

 

Output current

 

>1.0 A maximum. LED indication of over-current.

 

DC output switch

 

Sets output voltage level to zero when Off.

 

Output terminals

 

4mm terminals on 19mm (0.75") spacing. Duplicate terminals at rear (P versions only)

 

Output protection

 

Output will withstand up to 16 V forward voltage. Diode clamped for reverse voltages and 3 Amps reverse current.

 

Load regulation

 

<1% for 90% load change

 

Line regulation

 

<0.1% for 10% line change

 

Regulatory Approvals

 

CE-marked units meet: EN61010-1 and EN61326

 

 

 

 

1. General specifications apply for 5 to 4°C temperature range. Accuracy specifications apply for 18 to 28°C temperature range after 1 hour warm-up with no load and calibration at 23°C.

Typical specifications are determined by design and not guaranteed.

 

Note: Specifications are subject to change without notice.


Investigate our discount web specials
  Product Configurator  
DC Electronic Loads
DC Modular Supply
Discontinued Products
Product Support